In order to reduce the mismatch error, a direct current (DC) calibration method is introduced when the modulated microwave signal is measured. The microwave power is input to the left section of the power sensor, and the DC power is input to the right of the power sensor. Due to the existence of parasitic loss and electromagnetic coupling, the microwave power results in a mismatch error. However, the DC power does not have the mismatch error. So the DC power applied in the right section can calibrate the mismatch error of the microwave power in the left section. The calibration factor is measured at different modulation rates and modulation depths.The measurement results show that the carrier frequency is the major factor influencing the measurement results. After calibration, the carrier frequency and the modulation rate have little effect on the output voltage. The frequency response becomes relatively flat in the frequency range up to 20 GHz, and the sensitivity on average is enhanced by about 0.12 mV/dBm. Therefore, the DC calibration method has a certain reference value for the terminal-type microwave power sensor.