BACKGROUND:Paraquat (PQ) is a world-wide used herbicide and also a type of common poison for suicide and accidental poisoning. Numerous studies have proved that the concentration of serum PQ plays an important role in prognosis. Spectrophotometry, including common spectrophotometry and second-derivative spectrophotometry, is commonly used for PQ detection in primary hospitals. So far, lack of systematic research on the reliability of the method and the correlation between clinical features of patients with PQ poisoning and the test results has restricted the clinical use of spectrophotometry. This study aimed to evaluate the reliability and value of spectrophotometry in detecting the concentration of serum PQ. METHODS:The wavelengths for detecting the concentration of serum PQ by common and second-derivative spectrophotometry were determined. Second-derivative spectrophotometry was applied to detect the concentration of serum PQ. The linear range and precision for detection of PQ concentration by this method were confirmed. The concentration of serum PQ shown by second- derivative spectrophotometry and HPLC were compared in 8 patients with PQ poisoning. Altogether 21 patients with acute poisoning 4 hours after PQ ingestion treated in the period of October 2008 to September 2010 were retrospectively reviewed. The patients were divided into higher and lower than 1.8 μg/mL groups based on their concentrations of serum PQ measured by second-derivative spectrophotometry on admission. The severity of clinical manifestations between the two groups were analyzed with Student's t test or Fisher's exact test. RESULTS:The absorption peak of 257 nm could not be found when common spectrophotometry was used to detect the PQ concentration in serum. The calibration curve in the 0.4-8.0 μg/mL range for PQ concentration shown by second-derivative spectrophotometry obeyed Beer's law with r=0.996. The average recovery rates of PQ were within a range of 95.0% to 99.5%, relative standard deviation (RSD) was wit